Strain gauge load cells convert the kinetic energy of a force into a quantifiable electrical signal. However, this output signal is very small, in the order of millivolts. At the same time, the readout components that make those tiny load cell signals human-usable need a large, noise-free electrical input. Likewise, the process control systems that incorporate load cells also require stronger, cleaner signals than a load cell typically generates. The universal resolution to this disparity is the ever-important load cell amplifier (or other signal conditioner). These devices are the go-to for improving load cell signal quality and, in turn, weigh system accuracy.

This article explains the various forms of signal conditioning. But first, it gives a brief background of analog and digital signals and the process of converting one to the other. It covers this since analog-to-digital and digital-to-analog signal conversion are fundamental to signal conditioner technology. Finally, it concludes with a comparison table and buying guide to help the reader select the right product.
Analog and Digital Signals: How are they Related?
Most load and force measurement sensors or transducers generate an analog signal output. Therefore, signal processing has traditionally been analog. However, with the emergence of small computers, wireless Internet, and ever-evolving processing speed, real-time signal processing systems are increasingly digital and software-based.
Whereas analog processing remains faster, digital processing often delivers more accurate results. Digital signal conditioning also offers easier troubleshooting, adjustment, and tuning. Because each processing method has its advantages, signal conditioning systems generally incorporate both.
Analog-to-Digital Conversion
Analog-to-digital converters, or ADCs, appear in applications where altering or manipulating digital signals to get the desired output is simpler. Familiar examples are music recording or photography, where light or sound is converted to digital signals. The digital format gives tremendous flexibility in shaping these industries’ creations. Similarly, ADCs are useful in converting load cell measurements into a digital format.
ADCs sample the amplitude of a signal at discrete points in time to convert it to a digital signal representing the magnitude of the analog measurement. The arrows in Figure 4 below represent each moment data is sampled. The analog input, shown as the light green curve, is a constant stream along the time axis. The output samples are digitized into a stream of bytes.
Digital Signal Accuracy and Resolution
Signal sampling limits the allowable bandwidth, or range of frequencies that can be accurately represented by a digital signal. This is the “Nyquist limit,” and is equal to half the sampling frequency. That is, if a ADC samples at a rate of 120 samples per second, the highest input frequency that the ADC output can accurately represent is 60 Hz. Because the digital output is no longer a continuous signal, this sampling rate becomes extremely important. The higher the sampling rate, the better the digital signal approximates the analog one.
The number of digital bits used to represent the amplitude is also important. A bit is a binary digit, or logical value, that represents some unit of information. Each single bit represents one of two states (on-off or high-low). An analog-to-digital converter will typically use at least eight bits, with more bits increasing the converter’s signal accuracy or resolution. For example:
28 = 256 So the resolution of the 8 bit sample is 1 part of 256
212 = 4096 So the resolution of the 12 bit sample is 1 part of 4096
The more bits, the greater the dynamic range (how small or large a signal it can measure) and the lower the quantization error (the difference between the analog input and its digital approximation).
Signal Conditioners: What are they and What do they Achieve?
Signal conditioning is the process of converting the raw load cell electrical output to useful data. It can encompass one or more of these functions: amplification, attenuation, isolation, linearization, bridge completion, or other types of filtering.
When signal conditioning is part of a measuring system, its output will have these three qualities:
- It will be large enough to detect (through amplification),
- It will be free of noise/unwanted frequencies (through filtering), and
- It will be in the appropriate signaling format.
(1) Amplification: Analog and Digital
Most of the time, the raw output signal of a load cell is too weak to read without amplification, being of the order of 100mV or less. This makes the signal rather susceptible to ambient electrical noise (such as electromagnetic interference).
How does the load cell’s initial signal change with an instrumentation conditioner or amplifier? The load cell output is amplified from 0~36 mV to either 0-10V or 4-20mA. Amplifiers typically use DC power (24V) and can drive ~ 300Ω load cells connected directly. Many handle one or two devices alone, but can connect to more load cells at a time via junction boxes. (See Load Cell Summing: Junction Boxes, Signal Trim and Excitation Trim.)
Analog Amplifiers
Typically analog amplification happens through an operational amplifier, also known as an op-amp. Op-amps are electronic devices with two input terminals: the inverting and non-inverting ends. As Figure 2 shows, the raw signal connects to the inverting side with one resistor; the non-inverting side connects to ground. A feedback path connects from the output terminal, with a second resistor, to the inverting terminal.
The output signal is related to the input voltage through the equation below. Therefore the amount of amplification is based on the values of R1 and R2.
Operational Amplifier Relational Equation
\(V_{0}=\frac{R_{2}V_{i}}{R_{1}}\)
Instrumentation amplifiers, such as the ones used with load cells, require the amplification of extremely low-level signals. They may combine up to three op-amps to increase the output amplitude to the needed level.
Certain applications, like accelerometers, require the amplifier to have a high-frequency response, to prevent distortion.
Digital Amplifiers
Digital amplifiers, as the name implies, increase the magnitude of a digital signal rather than an analog one. When using these, the load cell’s analog output is converted to digital through an ADC before amplifying the signal. The digital amplifier then multiplies the sampled values by a fixed constant to amplify the signal.
(2) Attenuation
Attenuation is the opposite of amplification. Instead of increasing the signal amplitude, attenuation decreases it. This is necessary when the voltage is beyond the range of the analog-to-digital converter. Like digital amplification, a digital attenuator multiplies the digital signal by a fixed constant, mathematically changing the output’s amplitude.
(3) Isolation
Signals outside of the voltage range of an input device can cause equipment damage and hazards to operators. Isolation is common when there is a need to protect the measurement system from incidental voltage spikes. It also breaks ground loops, blocks surges, and rejects high voltages.
Various methods exist for isolating voltages, including creating physical barriers for signals, and using isolation amplifiers. The specific application, and the driving reasons for the need for isolation determine the method used.
(4) Signal Linearization
Signal linearization is the process of converting a transducer’s non-linear output into a linear signal to more accurately represent the measured value. This step is important when the relationship between a sensor’s input and output isn’t perfectly proportional. Devices that measure temperature, such as thermocouples, often need linearization to produce accurate readings across their full range.
Linearization can be done using analog signal conditioning, such as operational amplifiers, or through digital processing in software or microcontrollers.
(5) Bridge Completion
Bridge completion forms a four-resistor Wheatstone bridge when using a quarter or half-bridge sensor. The Tacuna Systems precision strain gauge amplifier/ conditioner provides this completion with high-precision resistors. This device detects small voltage changes across the active sensors connected to it.
(6) Signal Filtering
Signal filtering is a type of signal conditioning used to remove unwanted components, such as specific frequencies, from a raw signal. For example, if an operator wants to measure only the high-frequency portion of a load cell output, they can apply a filter to remove the low-frequency noise that might distort the results.
Signal filtering is possible both digitally and through analog circuits. The most common filters are high-pass, low-pass, band-pass, and notch filters. As their names suggest, these filters allow high, low, mid, or only high/low-frequency signals from the raw source signal to pass, while denying the rest. The range of permitted frequencies is called the “passband,” while the denied frequency range is called the “stopband.”
A passive filter uses only passive elements such as resistors, capacitors, inductors, and transformers. Therefore, it does not require an external power source. It has the added advantage of having a more stable output than an active filter.
An active filter, on the other hand, includes active components such as amplifiers and therefore requires a power source. Active filters can provide signal gain and better control over filtering characteristics, making them useful in many precision measurement systems.
How to Choose and Connect a Load Cell Amplifier or Conditioner
While choosing the appropriate load cell model for an application is important, selecting a signal conditioner is equally critical. The wrong amplifier or conditioner can distort the load cell measurements. Likewise, properly operating the load cell amplifier is critical to obtaining high-quality, accurate readings.
For best results in using signal conditioners, follow these three guidelines.
- Always follow the manufacturer’s operation manual,
- Ensure that operators receive proper training prior to installation,
- Contact your vendor should questions or concerns arise while operating products.
Tacuna Systems is readily available to answer any design or implementation questions regarding our amplifier and conditioner products. See our “Featured Customer Projects” section to see the scope of projects we’ve assisted with. (See especially, Next-Gen Transmissions for High-Performance Hybrid Cars, and Measuring Contact Forces on Soft Materials for Medical Robotics and Other Applications for test projects using the Tacuna Systems EMBSGB200 amplifier.)
Connecting and Installing a Load Cell Amplifier
The following is an example of connecting an ANYLOAD load cell amplifier:
- The input voltage should connect to 24V DC power and the ground terminal to the ground.
- The load cell connects to the amplifier through the positive(+) and negative(–) excitation terminals, and to the positive(+) and negative(–) output mV signal terminals.
- The output from the amplifier connects to either the current output for a signal of 4~20mA or a voltage output signal of 0~10V.
- To calibrate the conditioner, remove any load from the transducer and adjust the variable resistor to an output of 0V or 4mA. To calibrate the full span, place the entire load on the transducer and adjust its variable resistor to an output of 10V or 20mA.
The load measurement system will be fully functional and ready to operate. Re-calibrate the device frequently and whenever the output readings change, or the signal is abnormal. Most load cell signal conditioners should operate from 10°C – 40°C, or 15 – 100°F, and below 90% relative humidity. However, always review the product specs before operating in high and low-temperature environments.
Lastly, contact the supplier for setup assistance, calibration assistance, and troubleshooting.
Tacuna Systems Load Cell Amplifiers: Comparison and Buying Guide
This section provides a comparison table with each amplifier and signal conditioner product Tacuna Systems carries.
Device | Tacuna Systems EMBSGB200-M | Tacuna Systems EMBSGB200-C | Tacuna Systems EMBSGB200-X | ANYLOAD A1A-22 | ANYLOAD A1A-D25 | ANYLOAD A2A-D2 | ANYLOAD A2P-D2 |
---|---|---|---|---|---|---|---|
Load Cell Compatibility | Any strain gauge type | Any strain gauge type | Any strain gauge type | Any strain gauge type | Any strain gauge type | Any strain gauge type | Any strain gauge type |
Field Calibration Required? | No | No | No | Yes | Yes (9pt multi-section, non-linear calibration) | Yes | Yes |
Bridge Completion | Yes: 1/4, 1/2, 3/4, and full bridge | Yes: 1/4, 1/2, 3/4, and full bridge | Yes: 1/4, 1/2, 3/4, and full bridge | No | No | No | No |
Excitation voltage to LC | 5V | 5V | 5V | 12V | 5V | 5V or 12V (settable) | 5V or 12V (settable) |
Input Signal | 100mVDC | 100mVDC | 100mVDC | 0mV-30mV | 0mV-30mV | 0±36mV | 0±36mV |
Output Signal | 0-5V analog | 0-5V analog, 4-20mA | 0-5V analog or 12 bit digitized with ADC | 0-10V, 4-20mA | 0-10V | 0±10V, , 4-20mA | 0±10V, 4-20mA |
Noise Elimination Filter | Two low-pass (corner frequency ~70Hz with 350Ω bridge or ~100 Hz with 120Ω bridge) | Two low-pass (corner frequency ~70Hz with 350Ω bridge or ~100 Hz with 120Ω bridge) | One low pass antialiasing filter; corner frequency 200Hz | ~400Hz | 2.3Hz | 16Hz | |
Gain Setting | Discrete preset values selectable with DIP switch | Discrete preset values selectable with DIP switch | Programmable via RS232 | Potentiometer | Programmable via RS232 | Potentiometer (to adjust span) and combination of DIP switch and excitation voltage settings | Potentiometer (to adjust span) and combination of DIP switch and excitation voltage settings |
Offset Adjustment | Potentiometer | Potentiometer | Programmable via RS232 | Potentiometer | Programmable via RS232 | Potentiometer | Potentiometer |
Capacity | With junction box up to: 16 x 700Ω 8 x 350Ω | With junction box up to: 16 x 700Ω 8 x 350Ω | With junction box up to: 16 x 700Ω 8 x 350Ω | Single 350Ω or With junction box up to: 8 x 700Ω 4 x 350Ω | Single 350Ω or With junction box up to 4 x 350Ω or greater | Single 350Ω or With junction box up to: 8 x 700Ω 8 x 350Ω 4 x 350Ω | Single 350Ω or With junction box up to: 8 x 700Ω 8 x 350Ω 4 x 350Ω |
Overall Dimensions | 1in x 3in x 0.5in | 1in x 3in x 0.5in | 1in x 3in x 0.5in | 3.7in x 2.5in x 1.4in | 3.7in x 2.5in x 1.4in | 4.02in x 5.0in x 1.14in | 3.9in x 2.8in x 2.6in |
Enclosure Material | ABS (optional) | ABS (optional) | ABS (optional) | Aluminum Casting | Aluminum Casting | Aluminum Casting | PVC |
Enclosure Ingress Protection | IP40 | IP40 | IP40 | IP66 | IP66 | IP66 | IP40 |
Enclosure Installation | None – Tabletop or double-sided tape | None – Tabletop or double-sided tape | None – Tabletop or double-sided tape | Screw mounting holes within enclosure | Screw mounting holes within enclosure | GORE breather vent and capacitive screws | Screw mounting holes on flanges external to enclosure |
Summary
The raw signal from a load cell is often very small and can be affected by electrical noise or environmental factors. This makes it difficult for measurement interfaces to accurately convert the signal into a reliable force or weight reading.
Load cell signal conditioning, including amplification, filtering, and linearization, improves signal quality to ensure precise, consistent measurements. A properly conditioned output should be:
- Strong and stable enough for accurate processing by displays, controllers, or data systems
- In the correct format (analog or digital)
- Filtered to remove unwanted frequencies (noise-free)
Effective signal conditioning is essential for achieving accurate and dependable force measurement performance.
To see an interesting example of our Tacuna Systems EMBSGB200 Amplifier at work, see Next-Gen Transmissions for High-Performance Hybrid Cars in our “Featured Customer Projects” section of this Knowledge Base. Also, see How the EMBSGB200 Amplifier is Shaping Innovations in Technology and Industry for other examples of how our customers have used this versatile device.